ALP Autoclave Model CLG-32LDVP
Manufacturer: ALP
Origin: Japan
Contact Us
Mô tả
Thông số kĩ thuật
Phụ kiện
Tài liệu đính kèm
The ALP autoclave model CLG-32LDVP is the 4th model in the CLG Series. This is the most advanced line of Autoclave of the company using LCD screen with many functions and ensuring safety for users.
Besides, the pot has built-in Pre-Vacuum and Vacuum Drying functions to shorten the time and ensure the items (eg Clothes, tools) experiment) stay dry after sterilization.
Some line highlights Steamer sterilizer ALP model CLG-32LDVP:
- Temperature scale: 40 – 140°C
- Steam chamber volume: 62 Liters
- Maximum pressure: 0.270MPa
- Accessories: 2 sample baskets φ300×300mm
Features of ALP Autoclave Model CLG-32LDVP:
Pre-vacuum for safe sterilization
Air is removed from the chamber and loaded with a vacuum pump, removing air pockets to allow water vapor to penetrate every nook and cranny of the load. Thereby, helping to improve heat penetration, providing a safe sterilization.
Vacuum drying system
Dryers dry wet loads in a vacuum, the vacuum pump creates a place where moisture evaporates quickly. You can effectively dry even an empty tubular item and garment.
Integrate with 6 different programs depending on the load
You can choose six programs according to the load and application such as LIQUID “OPENED”, LIQUID “CLOSED”, SOLIDS, WASTE, DISSOLUTION, ADD DRYING. Furthermore, ten different memories are integrated with Model CLG-32LDVP allowing you to customize: temperature and time, drying or not, or cooling method during sterilization. .
Variety of applications
Model CLG-32LDVP can be used for sterilization:
Ferment jar
Dustproof uniform
Liquids in sealed vessels
Steam Release Pressure Pulse (SFPP)
CLG-32LDVP removes air from the load quickly by continuously alternating steam release and pressure pulse above atmospheric pressure . Optimal for sterilizing challenge loads.
Multi-program compatible
In addition to the five specified programs depending on the load such as LIQUID, SOLIDS and WASTE, etc. You can also set the temperature, sterilization time and cooling method to save with ten different memories.
Combines parts that are susceptible to damage over time
To avoid the running costs of accumulating due to parts that are prone to damage over time, CLG-32LDVP combines lid gasket with steam resistance, stainless steel element heater, etc. Thereby extending the life of these parts.
Qualified fast cooling fan
Cooling loads after operating cycles to shorten the time until unloading.
Liquid Crystal Interactive Display
The CLG-32LDVP‘s desktop settings are enlarged as shown below. One-touch switching between detail screen and zoom screen.
Data logging system (Optional)
You can measure temperature, pressure and time during operation into an optional software “ALP- LOGGER” by connecting the autoclave and a PC.
Data is displayed graphically “Temperature (Pressure) – Time”, which can be printed and exported to Microsoft Excel.
IQ/OQ profile conforms to GMP, GLP standards
Users can optionally purchase additional IQ/OQ documents of the Company.
Model | CLG-32LDVP |
Volume of autoclave | 62 Liters |
Used volume | 54 Liters |
Steam chamber size | Ø320×660 mm |
Outside size | 550x720x990 mm |
Weight | 109 kg |
Sterilizing temperature | 100 – 140°C |
Dry temperature | 60 – 150 °C |
Dissolution temperature | 40—99°C |
Warm up temperature | 40—60°C |
Time setting | 0 minutes – 48 hours 00 minutes |
Sterilization program | Can store 15 programs depending on user settings |
Maximum pressure | 0.270MPa |
Safety Features | Interlock lock, Electromechanical lock, Double check system for pot lid, Overpressure detection, Overheat detection, Water shortage prevention, Scene Probe failure alarm, Time difference warning, Data storage in case of power failure, Safety valve, Power off in case of leakage, over voltage or short circuit |
Pump | Watertight vacuum pump |
Vacuum money | Triple pulse vacuum system |
Vacuum drying | Vacuum pump removes moisture in chamber and dries |
Dry time | 0 minutes – 100 hours 00 minutes |
Power required | AC220/230/240V, single phase, 50/60Hz |
Standard accessory | 2 grid pattern baskets Ø300x300mm |
Supplied Accessories | |
Stainless steel mesh basket | |
Waste container | |
Optional accessories (optional) | |
Sterile temperature probe | |
Printer (directly mounted on Autoclave) | |
Cooling fan | |
Recorder | |
Other sterilization baskets |
Catalog Autoclave ALP - CLG series
Leave your questions, we will answer them right away