Microbiological Incubator with 2 compartments JSR JSGI-250DT
Contact Us
Tổng quan
JSR 2-compartment microbiological incubator model JSGI-250DT suitable for incubation needs of bacteria, fungi, cells cells and tissues on a temperature scale of +5°C to 70°C.
Some outstanding parameters of the series Microbiological incubator 2 compartments strong> JSR model JSGI-250DT:
+ Temperature scale: Ambient temperature +5 to 70°C
+ Volume: 500 Liters
+ Temperature accuracy: ±0.3°C
+ Temperature uniformity: ±1.5°C
Features of Microbiological incubator 2 compartments JSR model JSGI-250T:
Performance
– Temperature control in the range of Ambient Temperature +5 to 70°C.
– The air inside is forced by fan convection for high temperature uniformity and accuracy.
– The cabinet is designed with a circulating fan for superior temperature uniformity.
– Heating resistors are installed independently on the back of the cabinet to prevent contamination and easy to maintain.
Controllers
– Temperature control by microprocessor PID controller.
– User-friendly, backlit LCD display.
– UV coated Polycarbonate touch panel.
– There is a knob to change the setting value easily.
– Display resolution 0.1°C.
– Set time up to 99 hours 59 minutes and with warnings.
– Function to remember commonly used temperature and time values.
– Calibration settings at 3 points, 1 point offset.
– Function to restore setting data after power failure.
– Pt100Ω Class A probe.
– RS-232 computer connection port (option).
Structure
– The cabinet is made of durable eopxy coated metal.
– The cabinet is made of anti-corrosion 304 stainless steel, the inner corners are rounded for easy cleaning and leak prevention.
– Inner tempered glass door for easy viewing of samples.
– Port for connecting cables.
Convenient
– The exterior door has a self-closing magnet.
– Inner glass door with high density silicone gasket helps the door to be sealed airtight.
– The inner tray is removable and adjustable in height.
– The switch automatically disconnects the resistor and fan when the sampling door is opened.
Safety
– Door sensor with audible alarm when opening the door.
– Set maximum temperature limit, heating cut-off and audible warning.
– Stops heating when the temperature exceeds +2°C above the set temperature.
– Warning and power off when the temperature is too high.
– Warning and power off when the temperature is too low.
– Warning when connection to thermal probe is lost.
– Power cut off when voltage exceeds: fuse.
Leave your questions, we will answer them right away