Automatic Polarimeter Kruss Model P8000-TF
Manufacturer: Kruss
Origin: Germany
Contact Us
Mô tả
Thông số kĩ thuật
Phụ kiện
Tài liệu đính kèm
The Kruss Automatic Polarimeter Model P8000-TF is the first Model in the P8000-TF Series, allowing the entire measurement process to be semi-automatic or fully automatic and requires no user intervention, from sample supply to tube cleaning to tube drying.
Measurements can be user-independent and can work automatically. Thereby, maximize Efficiency and save time through automation options.
Some outstanding parameters of the Kruss Model P8000-TF Automatic Polarimeter:
Measurement unit: Optical rotation [°], Int. sugar scale [°Z], Concentration [g/100 ml], Spec. rotation, User-defined
Scale: ±90, ±259 °Z
Precision: ±0.003°, ±0.01 °Z
Resolution: 0.001°, 0.01°Z
Excellent Features:
– Measure the temperature directly inside the sample
– Precise, homogenous, fast circulating water bath
– Many choices with different sample tubes
– Quick coupling system for easy tube changes
– High precision on full scale
– Very short measurement time approx. 1 second
– Optical density measurement up to 3.0 OD
– Intuitive operation via touch screen
– Preset scales (opt. rotation, spec. rotation, int. sugar scale, concentration)
– Preset and definable scales with custom formula-based conversion values according to user requirements
– Chemical resistant material
– GMP / GLP Compliant, 21 CFR Part 11, Pharmaceutical (USP, BP, JP, Ph. Eur.), FDA, ISO, HACCP, OIML, ASTM, ICUMSA , NIST
Some pictures of Model P8000-TF:
Semi-automatic sampling | Autosampling |
|
|
Model P8000-TF’s semi-automatic system operated with the DS7070 peristaltic pump transfers the sample into the measuring chamber via a flow-through cell accessory , the measuring chamber is dried by the DS7060 dryer, the system is integrated with a 3/2-way valve so there is no need to change the hose when performing sample measurement or cleaning. | Model P8000-TF’s automated system operated with the DS7070 peristaltic pump transfers the sample into the measuring chamber via the flow-through cell accessory, The measuring chamber is dried by the DS7060 dryer, the AS80 and AS90 autosamplers allow the system to operate without user intervention. |
Model | P8000-TF |
Measurements | |
Measurement | Optical rotation [°]
Int. sugar scale [°Z] Concentration [g/100 ml] Spec. rotation User-defined |
Range | ±90°
±259 °Z |
Precision | ±0.003°
±0.01 °Z |
Resolution | 0.001°
0.01 °Z |
Measurement time | About 1 second |
Light source | 1 LED with filter |
Wavelength | 589nm |
Temperature parameters | |
Thermistor | Use PT31 circulating thermostat |
Thermistor scale | 8–35 °C |
Precision | ±0.2 °C |
Temperature method | Use PRT-E or PRT-T temperature probe |
Heat scale | 0–99.9 °C |
Resolution | 0.1 °C |
Precision | ±0.2 °C |
Measurement method | Unlimited number of measurement methods |
Adjustable | Automatic adjustment is available in the device menu |
Screen | 5.7”-TFT touch-screen, 640 x 480 Pixel |
Shell material | Aluminum cast, powder-coated |
Gateway | USB, RS-232, Ethernet |
Power | 110-250VAC 50/60Hz |
Optional Accessories
1/ Quartz standard tube
2/ Result printer, CBM910
3/ KrussLab software
Catalog Polarimeter Kruss
Leave your questions, we will answer them right away